Analytical Evaluation and Comparison of Dupont aca® Lactate Dehydrogenase-1 (LD1) Isoenzyme Assay Diagnostic Efficiency for Acute Myocardial Infarction Detection with Other LD1 Methods and aca® CK-MB

A Two-Site Study

Pennell C. Painter, Stuart Van Meter, Randal L. Dabbs, Gerald E. Clement

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

The purpose of this study was to examine the analtyical characteristics of a rapid new assay for lactate dehydrogenase 1 (LD1) isoenzyme on the Dupont aca® analyzer and the diagnostic efficiency of LD1 for detection of acute myocardial infarction (AMI) when used alone and with creatine kinase MB (CKMB). Total aca LD1 assay precision, with percent coefficients of variation (%CVs) of less than 3.2% from 56 U/L to 469 U/L LD1, across fifty assay days was excellent. Linearity was confirmed from 0 to 332 U/L and no detectable calibration drift was noted from 5 to 489 U/L over a ninety-day period. The aca LD1 results compared well with Roche Isomune-LD1, Abbott Spectrum A-Gent, Helena electrophoresis, and Beckman electrophoresis LD1 methods, giving r’s from 0.987 to 0.994, slopes from 0.94 to 1.65, and y-intercepts from -2.95 to 6.94 U/L. Examination of 450 ambulatory subjects, about equally distributed by sex, yielded a 43 ±14 U/L aca LD1 patient reference interval. Serum samples from 159 consecutive patients at the University of Tennessee Memorial Hospital and 96 patients at Allentown Hospital, submitted for AMI detection assistance, were assayed in a single-blind study for LD1 and ion-exchange CKMB by Dupont aca methods, which provide automated results in ten minutes whenever needed. The aca LD1 assay yielded a clinical sensitivity of 89% and specificity of 95% for AMI with a decision threshold of 120 U/L. The diagnostic effi ciency of the aca LD1 assay was 94% at 120 U/L, which equaled or exceeded that of the three comparative LD1 methods. The predictive value of positive (PV+) and negative (PV-) results on the first sample collected were 80% and 85% for aca LD1, 65% and 90% for aca ion-exchange CKMB, and 83% and 90% when both tests were combined. Significantly, the PV+ and PV- results when two or more samples were assayed was improved to 88% and 95% for aca LD1, relatively constant at 65% and 97% for aca ion- exchange CKMB, and dramatically improved to 95% and 100% when both CKMB and LD1 tests were combined. The results from these two medical centers show that the aca LD1 assay provides useful clinical information for AMI diagnosis when employed alone or in combination with aca CKMB. These results also suggest that LD1 should be included in biochemical evaluations for AMI to attain optimal predictive values of results.

Original languageEnglish (US)
Pages (from-to)585-595
Number of pages11
JournalAngiology
Volume45
Issue number7
DOIs
StatePublished - Jan 1 1994
Externally publishedYes

Fingerprint

Isoenzymes
Myocardial Infarction
MB Form Creatine Kinase
Ion Exchange
lactate dehydrogenase 1
L-Lactate Dehydrogenase
Electrophoresis
Single-Blind Method
Calibration

All Science Journal Classification (ASJC) codes

  • Cardiology and Cardiovascular Medicine

Cite this

@article{da72688210a74adbb7f046c4a810c698,
title = "Analytical Evaluation and Comparison of Dupont aca{\circledR} Lactate Dehydrogenase-1 (LD1) Isoenzyme Assay Diagnostic Efficiency for Acute Myocardial Infarction Detection with Other LD1 Methods and aca{\circledR} CK-MB: A Two-Site Study",
abstract = "The purpose of this study was to examine the analtyical characteristics of a rapid new assay for lactate dehydrogenase 1 (LD1) isoenzyme on the Dupont aca{\circledR} analyzer and the diagnostic efficiency of LD1 for detection of acute myocardial infarction (AMI) when used alone and with creatine kinase MB (CKMB). Total aca LD1 assay precision, with percent coefficients of variation ({\%}CVs) of less than 3.2{\%} from 56 U/L to 469 U/L LD1, across fifty assay days was excellent. Linearity was confirmed from 0 to 332 U/L and no detectable calibration drift was noted from 5 to 489 U/L over a ninety-day period. The aca LD1 results compared well with Roche Isomune-LD1, Abbott Spectrum A-Gent, Helena electrophoresis, and Beckman electrophoresis LD1 methods, giving r’s from 0.987 to 0.994, slopes from 0.94 to 1.65, and y-intercepts from -2.95 to 6.94 U/L. Examination of 450 ambulatory subjects, about equally distributed by sex, yielded a 43 ±14 U/L aca LD1 patient reference interval. Serum samples from 159 consecutive patients at the University of Tennessee Memorial Hospital and 96 patients at Allentown Hospital, submitted for AMI detection assistance, were assayed in a single-blind study for LD1 and ion-exchange CKMB by Dupont aca methods, which provide automated results in ten minutes whenever needed. The aca LD1 assay yielded a clinical sensitivity of 89{\%} and specificity of 95{\%} for AMI with a decision threshold of 120 U/L. The diagnostic effi ciency of the aca LD1 assay was 94{\%} at 120 U/L, which equaled or exceeded that of the three comparative LD1 methods. The predictive value of positive (PV+) and negative (PV-) results on the first sample collected were 80{\%} and 85{\%} for aca LD1, 65{\%} and 90{\%} for aca ion-exchange CKMB, and 83{\%} and 90{\%} when both tests were combined. Significantly, the PV+ and PV- results when two or more samples were assayed was improved to 88{\%} and 95{\%} for aca LD1, relatively constant at 65{\%} and 97{\%} for aca ion- exchange CKMB, and dramatically improved to 95{\%} and 100{\%} when both CKMB and LD1 tests were combined. The results from these two medical centers show that the aca LD1 assay provides useful clinical information for AMI diagnosis when employed alone or in combination with aca CKMB. These results also suggest that LD1 should be included in biochemical evaluations for AMI to attain optimal predictive values of results.",
author = "Painter, {Pennell C.} and {Van Meter}, Stuart and Dabbs, {Randal L.} and Clement, {Gerald E.}",
year = "1994",
month = "1",
day = "1",
doi = "10.1177/000331979404500701",
language = "English (US)",
volume = "45",
pages = "585--595",
journal = "Angiology",
issn = "0003-3197",
publisher = "SAGE Publications Inc.",
number = "7",

}

TY - JOUR

T1 - Analytical Evaluation and Comparison of Dupont aca® Lactate Dehydrogenase-1 (LD1) Isoenzyme Assay Diagnostic Efficiency for Acute Myocardial Infarction Detection with Other LD1 Methods and aca® CK-MB

T2 - A Two-Site Study

AU - Painter, Pennell C.

AU - Van Meter, Stuart

AU - Dabbs, Randal L.

AU - Clement, Gerald E.

PY - 1994/1/1

Y1 - 1994/1/1

N2 - The purpose of this study was to examine the analtyical characteristics of a rapid new assay for lactate dehydrogenase 1 (LD1) isoenzyme on the Dupont aca® analyzer and the diagnostic efficiency of LD1 for detection of acute myocardial infarction (AMI) when used alone and with creatine kinase MB (CKMB). Total aca LD1 assay precision, with percent coefficients of variation (%CVs) of less than 3.2% from 56 U/L to 469 U/L LD1, across fifty assay days was excellent. Linearity was confirmed from 0 to 332 U/L and no detectable calibration drift was noted from 5 to 489 U/L over a ninety-day period. The aca LD1 results compared well with Roche Isomune-LD1, Abbott Spectrum A-Gent, Helena electrophoresis, and Beckman electrophoresis LD1 methods, giving r’s from 0.987 to 0.994, slopes from 0.94 to 1.65, and y-intercepts from -2.95 to 6.94 U/L. Examination of 450 ambulatory subjects, about equally distributed by sex, yielded a 43 ±14 U/L aca LD1 patient reference interval. Serum samples from 159 consecutive patients at the University of Tennessee Memorial Hospital and 96 patients at Allentown Hospital, submitted for AMI detection assistance, were assayed in a single-blind study for LD1 and ion-exchange CKMB by Dupont aca methods, which provide automated results in ten minutes whenever needed. The aca LD1 assay yielded a clinical sensitivity of 89% and specificity of 95% for AMI with a decision threshold of 120 U/L. The diagnostic effi ciency of the aca LD1 assay was 94% at 120 U/L, which equaled or exceeded that of the three comparative LD1 methods. The predictive value of positive (PV+) and negative (PV-) results on the first sample collected were 80% and 85% for aca LD1, 65% and 90% for aca ion-exchange CKMB, and 83% and 90% when both tests were combined. Significantly, the PV+ and PV- results when two or more samples were assayed was improved to 88% and 95% for aca LD1, relatively constant at 65% and 97% for aca ion- exchange CKMB, and dramatically improved to 95% and 100% when both CKMB and LD1 tests were combined. The results from these two medical centers show that the aca LD1 assay provides useful clinical information for AMI diagnosis when employed alone or in combination with aca CKMB. These results also suggest that LD1 should be included in biochemical evaluations for AMI to attain optimal predictive values of results.

AB - The purpose of this study was to examine the analtyical characteristics of a rapid new assay for lactate dehydrogenase 1 (LD1) isoenzyme on the Dupont aca® analyzer and the diagnostic efficiency of LD1 for detection of acute myocardial infarction (AMI) when used alone and with creatine kinase MB (CKMB). Total aca LD1 assay precision, with percent coefficients of variation (%CVs) of less than 3.2% from 56 U/L to 469 U/L LD1, across fifty assay days was excellent. Linearity was confirmed from 0 to 332 U/L and no detectable calibration drift was noted from 5 to 489 U/L over a ninety-day period. The aca LD1 results compared well with Roche Isomune-LD1, Abbott Spectrum A-Gent, Helena electrophoresis, and Beckman electrophoresis LD1 methods, giving r’s from 0.987 to 0.994, slopes from 0.94 to 1.65, and y-intercepts from -2.95 to 6.94 U/L. Examination of 450 ambulatory subjects, about equally distributed by sex, yielded a 43 ±14 U/L aca LD1 patient reference interval. Serum samples from 159 consecutive patients at the University of Tennessee Memorial Hospital and 96 patients at Allentown Hospital, submitted for AMI detection assistance, were assayed in a single-blind study for LD1 and ion-exchange CKMB by Dupont aca methods, which provide automated results in ten minutes whenever needed. The aca LD1 assay yielded a clinical sensitivity of 89% and specificity of 95% for AMI with a decision threshold of 120 U/L. The diagnostic effi ciency of the aca LD1 assay was 94% at 120 U/L, which equaled or exceeded that of the three comparative LD1 methods. The predictive value of positive (PV+) and negative (PV-) results on the first sample collected were 80% and 85% for aca LD1, 65% and 90% for aca ion-exchange CKMB, and 83% and 90% when both tests were combined. Significantly, the PV+ and PV- results when two or more samples were assayed was improved to 88% and 95% for aca LD1, relatively constant at 65% and 97% for aca ion- exchange CKMB, and dramatically improved to 95% and 100% when both CKMB and LD1 tests were combined. The results from these two medical centers show that the aca LD1 assay provides useful clinical information for AMI diagnosis when employed alone or in combination with aca CKMB. These results also suggest that LD1 should be included in biochemical evaluations for AMI to attain optimal predictive values of results.

UR - http://www.scopus.com/inward/record.url?scp=0028361289&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0028361289&partnerID=8YFLogxK

U2 - 10.1177/000331979404500701

DO - 10.1177/000331979404500701

M3 - Article

VL - 45

SP - 585

EP - 595

JO - Angiology

JF - Angiology

SN - 0003-3197

IS - 7

ER -