Abstract

An experiment was conducted to determine the influence of anterior cervical plates on multi-level strut-graft loading mechanisms in vitro. Four fresh cadaveric cervical spines were harvested and radiographed and were biomechanically tested in four sequential conditions: harvested (H), (C4-C6) corpectomy (C), strut-grafted (SG), and SG with anterior cervical plate (SGP). Following corpectomy testing, the spine was decompressed and a force sensing strut graft (FSSG) was inserted. The plated spine included an Orion plate attached between C3-C7. Attachment of plate caused an immediate increase in the FSSG load. During flexion of the SGP spine, the FSSG load decreased. However, extension increased the FSSG load.

Original languageEnglish (US)
Number of pages1
StatePublished - Jan 1 1998
EventProceedings of the 1998 17th Southern Biomedical Engineering Conference - San Antonio, TX, USA
Duration: Feb 6 1998Feb 8 1998

Other

OtherProceedings of the 1998 17th Southern Biomedical Engineering Conference
CitySan Antonio, TX, USA
Period2/6/982/8/98

Fingerprint

Struts
Plating
Grafts
Mechanics
Testing

All Science Journal Classification (ASJC) codes

  • Engineering(all)

Cite this

Diangelo, D., Foley, K., Jansen, T. H., & Vossel, K. A. (1998). Anterior cervical plating reverses the loading mechanics of multi-level strut grafts during flexion and extension. Paper presented at Proceedings of the 1998 17th Southern Biomedical Engineering Conference, San Antonio, TX, USA, .

Anterior cervical plating reverses the loading mechanics of multi-level strut grafts during flexion and extension. / Diangelo, Denis; Foley, Kevin; Jansen, Thomas H.; Vossel, Keith A.

1998. Paper presented at Proceedings of the 1998 17th Southern Biomedical Engineering Conference, San Antonio, TX, USA, .

Research output: Contribution to conferencePaper

Diangelo, D, Foley, K, Jansen, TH & Vossel, KA 1998, 'Anterior cervical plating reverses the loading mechanics of multi-level strut grafts during flexion and extension' Paper presented at Proceedings of the 1998 17th Southern Biomedical Engineering Conference, San Antonio, TX, USA, 2/6/98 - 2/8/98, .
Diangelo D, Foley K, Jansen TH, Vossel KA. Anterior cervical plating reverses the loading mechanics of multi-level strut grafts during flexion and extension. 1998. Paper presented at Proceedings of the 1998 17th Southern Biomedical Engineering Conference, San Antonio, TX, USA, .
Diangelo, Denis ; Foley, Kevin ; Jansen, Thomas H. ; Vossel, Keith A. / Anterior cervical plating reverses the loading mechanics of multi-level strut grafts during flexion and extension. Paper presented at Proceedings of the 1998 17th Southern Biomedical Engineering Conference, San Antonio, TX, USA, .1 p.
@conference{965c5d086861496f8e158593edb2bc3a,
title = "Anterior cervical plating reverses the loading mechanics of multi-level strut grafts during flexion and extension",
abstract = "An experiment was conducted to determine the influence of anterior cervical plates on multi-level strut-graft loading mechanisms in vitro. Four fresh cadaveric cervical spines were harvested and radiographed and were biomechanically tested in four sequential conditions: harvested (H), (C4-C6) corpectomy (C), strut-grafted (SG), and SG with anterior cervical plate (SGP). Following corpectomy testing, the spine was decompressed and a force sensing strut graft (FSSG) was inserted. The plated spine included an Orion plate attached between C3-C7. Attachment of plate caused an immediate increase in the FSSG load. During flexion of the SGP spine, the FSSG load decreased. However, extension increased the FSSG load.",
author = "Denis Diangelo and Kevin Foley and Jansen, {Thomas H.} and Vossel, {Keith A.}",
year = "1998",
month = "1",
day = "1",
language = "English (US)",
note = "Proceedings of the 1998 17th Southern Biomedical Engineering Conference ; Conference date: 06-02-1998 Through 08-02-1998",

}

TY - CONF

T1 - Anterior cervical plating reverses the loading mechanics of multi-level strut grafts during flexion and extension

AU - Diangelo, Denis

AU - Foley, Kevin

AU - Jansen, Thomas H.

AU - Vossel, Keith A.

PY - 1998/1/1

Y1 - 1998/1/1

N2 - An experiment was conducted to determine the influence of anterior cervical plates on multi-level strut-graft loading mechanisms in vitro. Four fresh cadaveric cervical spines were harvested and radiographed and were biomechanically tested in four sequential conditions: harvested (H), (C4-C6) corpectomy (C), strut-grafted (SG), and SG with anterior cervical plate (SGP). Following corpectomy testing, the spine was decompressed and a force sensing strut graft (FSSG) was inserted. The plated spine included an Orion plate attached between C3-C7. Attachment of plate caused an immediate increase in the FSSG load. During flexion of the SGP spine, the FSSG load decreased. However, extension increased the FSSG load.

AB - An experiment was conducted to determine the influence of anterior cervical plates on multi-level strut-graft loading mechanisms in vitro. Four fresh cadaveric cervical spines were harvested and radiographed and were biomechanically tested in four sequential conditions: harvested (H), (C4-C6) corpectomy (C), strut-grafted (SG), and SG with anterior cervical plate (SGP). Following corpectomy testing, the spine was decompressed and a force sensing strut graft (FSSG) was inserted. The plated spine included an Orion plate attached between C3-C7. Attachment of plate caused an immediate increase in the FSSG load. During flexion of the SGP spine, the FSSG load decreased. However, extension increased the FSSG load.

UR - http://www.scopus.com/inward/record.url?scp=0031653803&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0031653803&partnerID=8YFLogxK

M3 - Paper

ER -