A patient with muscle cramps was referred for nerve conduction studies and electromyography. Her study demonstrated bursts of spontaneous electromyography activity, which waxed and waned in amplitude with a "dive bomber"-like sound. The abnormal finding was found to be an artifact from an implanted sacral nerve stimulator for bladder incontinence.

Original languageEnglish (US)
Pages (from-to)77-78
Number of pages2
JournalJournal of Clinical Neuromuscular Disease
Volume11
Issue number2
DOIs
StatePublished - Dec 1 2009

Fingerprint

Electromyography
Artifacts
Muscle Cramp
Neural Conduction
Urinary Bladder

All Science Journal Classification (ASJC) codes

  • Clinical Neurology
  • Neurology

Cite this

Artifact on electromyography produced by an implanted Sacral nerve stimulator. / O'Carroll, Peter; Bertorini, Tulio; MacCarino, Sue.

In: Journal of Clinical Neuromuscular Disease, Vol. 11, No. 2, 01.12.2009, p. 77-78.

Research output: Contribution to journalArticle

@article{f41d3a9b6dc04249a1d0d14633953e0f,
title = "Artifact on electromyography produced by an implanted Sacral nerve stimulator",
abstract = "A patient with muscle cramps was referred for nerve conduction studies and electromyography. Her study demonstrated bursts of spontaneous electromyography activity, which waxed and waned in amplitude with a {"}dive bomber{"}-like sound. The abnormal finding was found to be an artifact from an implanted sacral nerve stimulator for bladder incontinence.",
author = "Peter O'Carroll and Tulio Bertorini and Sue MacCarino",
year = "2009",
month = "12",
day = "1",
doi = "10.1097/CND.0b013e3181b40d37",
language = "English (US)",
volume = "11",
pages = "77--78",
journal = "Journal of Clinical Neuromuscular Disease",
issn = "1522-0443",
publisher = "Lippincott Williams and Wilkins",
number = "2",

}

TY - JOUR

T1 - Artifact on electromyography produced by an implanted Sacral nerve stimulator

AU - O'Carroll, Peter

AU - Bertorini, Tulio

AU - MacCarino, Sue

PY - 2009/12/1

Y1 - 2009/12/1

N2 - A patient with muscle cramps was referred for nerve conduction studies and electromyography. Her study demonstrated bursts of spontaneous electromyography activity, which waxed and waned in amplitude with a "dive bomber"-like sound. The abnormal finding was found to be an artifact from an implanted sacral nerve stimulator for bladder incontinence.

AB - A patient with muscle cramps was referred for nerve conduction studies and electromyography. Her study demonstrated bursts of spontaneous electromyography activity, which waxed and waned in amplitude with a "dive bomber"-like sound. The abnormal finding was found to be an artifact from an implanted sacral nerve stimulator for bladder incontinence.

UR - http://www.scopus.com/inward/record.url?scp=75149192735&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=75149192735&partnerID=8YFLogxK

U2 - 10.1097/CND.0b013e3181b40d37

DO - 10.1097/CND.0b013e3181b40d37

M3 - Article

VL - 11

SP - 77

EP - 78

JO - Journal of Clinical Neuromuscular Disease

JF - Journal of Clinical Neuromuscular Disease

SN - 1522-0443

IS - 2

ER -