Intraluminal stented grafts are currently the focus of much attention and research. Increasing use of balloon-expandable stents in these devices has resulted in a spectrum of deployment associated complications and new technical challenges for surgeons. The various techniques used to overcome problems during intravascular stented graft deployment are reported. Deployment complications occurred in 3 of 19 adult hound dogs subjected to endovascular stented graft placement. The stented grafts consisted of a balloon expandable stent sewn to a 6-cm-long,10.mm-diameter, thin-walled polytetrafluoroethylene graft. Reengaging a nondeployed stented graft that slid off an expansion balloon, managing a partially deployed stented graft with inability to disengage a broken angioplasty balloon, and repositioning a free stent within the aorta are described. Treatment of stented graft complications was associated with no morbidity or mortality and minimal blood loss and added only fifteen to thirty minutes of procedure time. All situations were managed successfully without the need for an open retrieval procedure. Deployment complications associated with intravascular stented grafts may be successfully managed with endovascular techniques using the existing graft delivery system and additional small-diameter balloon catheters.

Original languageEnglish (US)
Pages (from-to)293-299
Number of pages7
JournalVascular and Endovascular Surgery
Volume30
Issue number4
DOIs
StatePublished - Jan 1 1996

Fingerprint

Stents
Transplants
Endovascular Procedures
Balloon Angioplasty
Polytetrafluoroethylene
Aorta
Catheters
Dogs
Morbidity
Equipment and Supplies
Mortality
Research

All Science Journal Classification (ASJC) codes

  • Surgery
  • Cardiology and Cardiovascular Medicine

Cite this

Balloon-Expandable Stent Complications During Endovascular Grafting. / Ombrellaro, Mark P.; Stevens, Scott; Kerstetter, Kyle; Freeman, Michael; Goldman, Mitchell.

In: Vascular and Endovascular Surgery, Vol. 30, No. 4, 01.01.1996, p. 293-299.

Research output: Contribution to journalArticle

@article{7cb40e454f5f4449ad098bc891223a8c,
title = "Balloon-Expandable Stent Complications During Endovascular Grafting",
abstract = "Intraluminal stented grafts are currently the focus of much attention and research. Increasing use of balloon-expandable stents in these devices has resulted in a spectrum of deployment associated complications and new technical challenges for surgeons. The various techniques used to overcome problems during intravascular stented graft deployment are reported. Deployment complications occurred in 3 of 19 adult hound dogs subjected to endovascular stented graft placement. The stented grafts consisted of a balloon expandable stent sewn to a 6-cm-long,10.mm-diameter, thin-walled polytetrafluoroethylene graft. Reengaging a nondeployed stented graft that slid off an expansion balloon, managing a partially deployed stented graft with inability to disengage a broken angioplasty balloon, and repositioning a free stent within the aorta are described. Treatment of stented graft complications was associated with no morbidity or mortality and minimal blood loss and added only fifteen to thirty minutes of procedure time. All situations were managed successfully without the need for an open retrieval procedure. Deployment complications associated with intravascular stented grafts may be successfully managed with endovascular techniques using the existing graft delivery system and additional small-diameter balloon catheters.",
author = "Ombrellaro, {Mark P.} and Scott Stevens and Kyle Kerstetter and Michael Freeman and Mitchell Goldman",
year = "1996",
month = "1",
day = "1",
doi = "10.1177/153857449603000405",
language = "English (US)",
volume = "30",
pages = "293--299",
journal = "Vascular and Endovascular Surgery",
issn = "1538-5744",
publisher = "SAGE Publications Inc.",
number = "4",

}

TY - JOUR

T1 - Balloon-Expandable Stent Complications During Endovascular Grafting

AU - Ombrellaro, Mark P.

AU - Stevens, Scott

AU - Kerstetter, Kyle

AU - Freeman, Michael

AU - Goldman, Mitchell

PY - 1996/1/1

Y1 - 1996/1/1

N2 - Intraluminal stented grafts are currently the focus of much attention and research. Increasing use of balloon-expandable stents in these devices has resulted in a spectrum of deployment associated complications and new technical challenges for surgeons. The various techniques used to overcome problems during intravascular stented graft deployment are reported. Deployment complications occurred in 3 of 19 adult hound dogs subjected to endovascular stented graft placement. The stented grafts consisted of a balloon expandable stent sewn to a 6-cm-long,10.mm-diameter, thin-walled polytetrafluoroethylene graft. Reengaging a nondeployed stented graft that slid off an expansion balloon, managing a partially deployed stented graft with inability to disengage a broken angioplasty balloon, and repositioning a free stent within the aorta are described. Treatment of stented graft complications was associated with no morbidity or mortality and minimal blood loss and added only fifteen to thirty minutes of procedure time. All situations were managed successfully without the need for an open retrieval procedure. Deployment complications associated with intravascular stented grafts may be successfully managed with endovascular techniques using the existing graft delivery system and additional small-diameter balloon catheters.

AB - Intraluminal stented grafts are currently the focus of much attention and research. Increasing use of balloon-expandable stents in these devices has resulted in a spectrum of deployment associated complications and new technical challenges for surgeons. The various techniques used to overcome problems during intravascular stented graft deployment are reported. Deployment complications occurred in 3 of 19 adult hound dogs subjected to endovascular stented graft placement. The stented grafts consisted of a balloon expandable stent sewn to a 6-cm-long,10.mm-diameter, thin-walled polytetrafluoroethylene graft. Reengaging a nondeployed stented graft that slid off an expansion balloon, managing a partially deployed stented graft with inability to disengage a broken angioplasty balloon, and repositioning a free stent within the aorta are described. Treatment of stented graft complications was associated with no morbidity or mortality and minimal blood loss and added only fifteen to thirty minutes of procedure time. All situations were managed successfully without the need for an open retrieval procedure. Deployment complications associated with intravascular stented grafts may be successfully managed with endovascular techniques using the existing graft delivery system and additional small-diameter balloon catheters.

UR - http://www.scopus.com/inward/record.url?scp=0029971054&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0029971054&partnerID=8YFLogxK

U2 - 10.1177/153857449603000405

DO - 10.1177/153857449603000405

M3 - Article

VL - 30

SP - 293

EP - 299

JO - Vascular and Endovascular Surgery

JF - Vascular and Endovascular Surgery

SN - 1538-5744

IS - 4

ER -