Fine tilt tuning of a laparoscopic camera by local magnetic actuation

Two-port nephrectomy experience on human cadavers

Massimiliano Simi, Ryan Pickens, Arianna Menciassi, S. Duke Herrell, Pietro Valdastri

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

Background. The magnetic surgical camera is an emerging technology having the potential to improve visualization without taking up port site space. However, tilting the point of view downward/upward can be done only by constantly applying a pressure on the abdomen. This study aims to test the hypothesis that the novel concept of local magnetic actuation (LMA) is able to increase the tilt range available for a magnetic camera without the need for deforming the abdominal wall. The hypothesis that 2-port laparoscopic nephrectomy in fresh tissue human cadavers could be performed by using the LMA camera is also tested. Methods. First, the 2 cameras were separately inserted, anchored, and moved inside the inflated abdomen. Tilting angles were quantified by image analysis while intra-abdominal pressure changes were monitored. Then, 5 two-port nephrectomies were performed by using the LMA camera while collecting quantitative outcomes. Results. The magnetic camera required a constant pressure on the magnetic handle to achieve an average ±20 tilt from the horizontal position, with an average of 7 mm Hg loss of intra-abdominal pressure. The LMA camera allowed for 75 of tilt from the horizontal position with a resolution of ±1 , without any need to deform the abdomen. All the nephrectomies were completed successfully within an average time of 11 minutes. Conclusion. LMA is an effective strategy to provide magnetic cameras with wide-range and high-resolution vertical motion without the need to deform the abdominal wall.

Original languageEnglish (US)
Pages (from-to)385-394
Number of pages10
JournalSurgical Innovation
Volume20
Issue number4
DOIs
StatePublished - Aug 1 2013
Externally publishedYes

Fingerprint

Nephrectomy
Cadaver
Abdomen
Pressure
Abdominal Wall
Technology

All Science Journal Classification (ASJC) codes

  • Surgery

Cite this

Fine tilt tuning of a laparoscopic camera by local magnetic actuation : Two-port nephrectomy experience on human cadavers. / Simi, Massimiliano; Pickens, Ryan; Menciassi, Arianna; Herrell, S. Duke; Valdastri, Pietro.

In: Surgical Innovation, Vol. 20, No. 4, 01.08.2013, p. 385-394.

Research output: Contribution to journalArticle

Simi, Massimiliano ; Pickens, Ryan ; Menciassi, Arianna ; Herrell, S. Duke ; Valdastri, Pietro. / Fine tilt tuning of a laparoscopic camera by local magnetic actuation : Two-port nephrectomy experience on human cadavers. In: Surgical Innovation. 2013 ; Vol. 20, No. 4. pp. 385-394.
@article{9a036f963185483b88fa1103a662885c,
title = "Fine tilt tuning of a laparoscopic camera by local magnetic actuation: Two-port nephrectomy experience on human cadavers",
abstract = "Background. The magnetic surgical camera is an emerging technology having the potential to improve visualization without taking up port site space. However, tilting the point of view downward/upward can be done only by constantly applying a pressure on the abdomen. This study aims to test the hypothesis that the novel concept of local magnetic actuation (LMA) is able to increase the tilt range available for a magnetic camera without the need for deforming the abdominal wall. The hypothesis that 2-port laparoscopic nephrectomy in fresh tissue human cadavers could be performed by using the LMA camera is also tested. Methods. First, the 2 cameras were separately inserted, anchored, and moved inside the inflated abdomen. Tilting angles were quantified by image analysis while intra-abdominal pressure changes were monitored. Then, 5 two-port nephrectomies were performed by using the LMA camera while collecting quantitative outcomes. Results. The magnetic camera required a constant pressure on the magnetic handle to achieve an average ±20 tilt from the horizontal position, with an average of 7 mm Hg loss of intra-abdominal pressure. The LMA camera allowed for 75 of tilt from the horizontal position with a resolution of ±1 , without any need to deform the abdomen. All the nephrectomies were completed successfully within an average time of 11 minutes. Conclusion. LMA is an effective strategy to provide magnetic cameras with wide-range and high-resolution vertical motion without the need to deform the abdominal wall.",
author = "Massimiliano Simi and Ryan Pickens and Arianna Menciassi and Herrell, {S. Duke} and Pietro Valdastri",
year = "2013",
month = "8",
day = "1",
doi = "10.1177/1553350612462458",
language = "English (US)",
volume = "20",
pages = "385--394",
journal = "Surgical Innovation",
issn = "1553-3506",
publisher = "SAGE Publications Inc.",
number = "4",

}

TY - JOUR

T1 - Fine tilt tuning of a laparoscopic camera by local magnetic actuation

T2 - Two-port nephrectomy experience on human cadavers

AU - Simi, Massimiliano

AU - Pickens, Ryan

AU - Menciassi, Arianna

AU - Herrell, S. Duke

AU - Valdastri, Pietro

PY - 2013/8/1

Y1 - 2013/8/1

N2 - Background. The magnetic surgical camera is an emerging technology having the potential to improve visualization without taking up port site space. However, tilting the point of view downward/upward can be done only by constantly applying a pressure on the abdomen. This study aims to test the hypothesis that the novel concept of local magnetic actuation (LMA) is able to increase the tilt range available for a magnetic camera without the need for deforming the abdominal wall. The hypothesis that 2-port laparoscopic nephrectomy in fresh tissue human cadavers could be performed by using the LMA camera is also tested. Methods. First, the 2 cameras were separately inserted, anchored, and moved inside the inflated abdomen. Tilting angles were quantified by image analysis while intra-abdominal pressure changes were monitored. Then, 5 two-port nephrectomies were performed by using the LMA camera while collecting quantitative outcomes. Results. The magnetic camera required a constant pressure on the magnetic handle to achieve an average ±20 tilt from the horizontal position, with an average of 7 mm Hg loss of intra-abdominal pressure. The LMA camera allowed for 75 of tilt from the horizontal position with a resolution of ±1 , without any need to deform the abdomen. All the nephrectomies were completed successfully within an average time of 11 minutes. Conclusion. LMA is an effective strategy to provide magnetic cameras with wide-range and high-resolution vertical motion without the need to deform the abdominal wall.

AB - Background. The magnetic surgical camera is an emerging technology having the potential to improve visualization without taking up port site space. However, tilting the point of view downward/upward can be done only by constantly applying a pressure on the abdomen. This study aims to test the hypothesis that the novel concept of local magnetic actuation (LMA) is able to increase the tilt range available for a magnetic camera without the need for deforming the abdominal wall. The hypothesis that 2-port laparoscopic nephrectomy in fresh tissue human cadavers could be performed by using the LMA camera is also tested. Methods. First, the 2 cameras were separately inserted, anchored, and moved inside the inflated abdomen. Tilting angles were quantified by image analysis while intra-abdominal pressure changes were monitored. Then, 5 two-port nephrectomies were performed by using the LMA camera while collecting quantitative outcomes. Results. The magnetic camera required a constant pressure on the magnetic handle to achieve an average ±20 tilt from the horizontal position, with an average of 7 mm Hg loss of intra-abdominal pressure. The LMA camera allowed for 75 of tilt from the horizontal position with a resolution of ±1 , without any need to deform the abdomen. All the nephrectomies were completed successfully within an average time of 11 minutes. Conclusion. LMA is an effective strategy to provide magnetic cameras with wide-range and high-resolution vertical motion without the need to deform the abdominal wall.

UR - http://www.scopus.com/inward/record.url?scp=84880283829&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84880283829&partnerID=8YFLogxK

U2 - 10.1177/1553350612462458

DO - 10.1177/1553350612462458

M3 - Article

VL - 20

SP - 385

EP - 394

JO - Surgical Innovation

JF - Surgical Innovation

SN - 1553-3506

IS - 4