TH‐C‐351‐01

Design of the Washington University Small Animal Conformal Micro Irradiator

Enrique Izaguirre, X. Diao, S. Stojadinovic, S. Mutic, D. Khullar, J. Deasy, P. Parikh, A. Hope, P. Grigsby, D. A. Low

Research output: Contribution to journalArticle

Abstract

Purpose: To design a conformal small animal micro radiation therapy (microRT) instrument, consisting of a cone beam microCT subsystem for submillimeter low dose structural imaging and image guided radiotherapy and an orthovoltage conformal micro irradiator with high radiation dose rate for high throughput conformal irradiation. Method and Materials: The microCT subsystem is based on an 80kVp micro‐focus x‐ray source with 75×75 um2 focal spot and a flat panel amorphous silicon detector with 1024×1024 pixels. The microRT subsystem design utilizes a 320kVp orthovoltage source with dual focus spots (0.4×0.4 mm2 at 800W and 1×1 mm2 at 1800 W). The orthovoltage beam is collimated using two orthogonal jaws and exchangeable apertures. The resolution of the cone‐beam micro CT and the therapy beam spatial precision was determined using a numerical model. The microCT radiation dose, the orthovoltage source spectral output, and dose rate were evaluated using a mouse digital phantom (MOBY) and a pencil beam algorithm. Results: microCT reconstructed tomographic data with a resolution of 125 μm is achievable using 128 projections and a maximum radiation dose of 2cGy. Automatic animal positioning and handling is performed within a precision of 100 μm. The treatment beam can be aimed at different latitude and longitude angles in steps of 2 arc min. and translated at 50μm steps (x,y,z). The beam cross section can be modulated with submillimeter precision using steps of 50μm. A radiation dose rate of 40 Gy/min is delivered when the system is operated at an average half‐value layer of 4.6 mm Cu and a maximum beam penumbra of 0.35 mm. Conclusion: We designed and numerically evaluated the performance of a microRT system integrated with a microCT for conformal radiotherapy of murine animal models. This work supported in part by NIH grant CA108677.

Original languageEnglish (US)
Number of pages1
JournalMedical physics
Volume35
Issue number6
DOIs
StatePublished - Jan 1 2008

Fingerprint

X-Ray Microtomography
Radiation
Radiotherapy
Image-Guided Radiotherapy
Conformal Radiotherapy
Organized Financing
Silicon
Jaw
Animal Models
X-Rays
Therapeutics

All Science Journal Classification (ASJC) codes

  • Biophysics
  • Radiology Nuclear Medicine and imaging

Cite this

TH‐C‐351‐01 : Design of the Washington University Small Animal Conformal Micro Irradiator. / Izaguirre, Enrique; Diao, X.; Stojadinovic, S.; Mutic, S.; Khullar, D.; Deasy, J.; Parikh, P.; Hope, A.; Grigsby, P.; Low, D. A.

In: Medical physics, Vol. 35, No. 6, 01.01.2008.

Research output: Contribution to journalArticle

Izaguirre, E, Diao, X, Stojadinovic, S, Mutic, S, Khullar, D, Deasy, J, Parikh, P, Hope, A, Grigsby, P & Low, DA 2008, 'TH‐C‐351‐01: Design of the Washington University Small Animal Conformal Micro Irradiator', Medical physics, vol. 35, no. 6. https://doi.org/10.1118/1.2962857
Izaguirre, Enrique ; Diao, X. ; Stojadinovic, S. ; Mutic, S. ; Khullar, D. ; Deasy, J. ; Parikh, P. ; Hope, A. ; Grigsby, P. ; Low, D. A. / TH‐C‐351‐01 : Design of the Washington University Small Animal Conformal Micro Irradiator. In: Medical physics. 2008 ; Vol. 35, No. 6.
@article{115b460c440b496cb04746713da85b64,
title = "TH‐C‐351‐01: Design of the Washington University Small Animal Conformal Micro Irradiator",
abstract = "Purpose: To design a conformal small animal micro radiation therapy (microRT) instrument, consisting of a cone beam microCT subsystem for submillimeter low dose structural imaging and image guided radiotherapy and an orthovoltage conformal micro irradiator with high radiation dose rate for high throughput conformal irradiation. Method and Materials: The microCT subsystem is based on an 80kVp micro‐focus x‐ray source with 75×75 um2 focal spot and a flat panel amorphous silicon detector with 1024×1024 pixels. The microRT subsystem design utilizes a 320kVp orthovoltage source with dual focus spots (0.4×0.4 mm2 at 800W and 1×1 mm2 at 1800 W). The orthovoltage beam is collimated using two orthogonal jaws and exchangeable apertures. The resolution of the cone‐beam micro CT and the therapy beam spatial precision was determined using a numerical model. The microCT radiation dose, the orthovoltage source spectral output, and dose rate were evaluated using a mouse digital phantom (MOBY) and a pencil beam algorithm. Results: microCT reconstructed tomographic data with a resolution of 125 μm is achievable using 128 projections and a maximum radiation dose of 2cGy. Automatic animal positioning and handling is performed within a precision of 100 μm. The treatment beam can be aimed at different latitude and longitude angles in steps of 2 arc min. and translated at 50μm steps (x,y,z). The beam cross section can be modulated with submillimeter precision using steps of 50μm. A radiation dose rate of 40 Gy/min is delivered when the system is operated at an average half‐value layer of 4.6 mm Cu and a maximum beam penumbra of 0.35 mm. Conclusion: We designed and numerically evaluated the performance of a microRT system integrated with a microCT for conformal radiotherapy of murine animal models. This work supported in part by NIH grant CA108677.",
author = "Enrique Izaguirre and X. Diao and S. Stojadinovic and S. Mutic and D. Khullar and J. Deasy and P. Parikh and A. Hope and P. Grigsby and Low, {D. A.}",
year = "2008",
month = "1",
day = "1",
doi = "10.1118/1.2962857",
language = "English (US)",
volume = "35",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - TH‐C‐351‐01

T2 - Design of the Washington University Small Animal Conformal Micro Irradiator

AU - Izaguirre, Enrique

AU - Diao, X.

AU - Stojadinovic, S.

AU - Mutic, S.

AU - Khullar, D.

AU - Deasy, J.

AU - Parikh, P.

AU - Hope, A.

AU - Grigsby, P.

AU - Low, D. A.

PY - 2008/1/1

Y1 - 2008/1/1

N2 - Purpose: To design a conformal small animal micro radiation therapy (microRT) instrument, consisting of a cone beam microCT subsystem for submillimeter low dose structural imaging and image guided radiotherapy and an orthovoltage conformal micro irradiator with high radiation dose rate for high throughput conformal irradiation. Method and Materials: The microCT subsystem is based on an 80kVp micro‐focus x‐ray source with 75×75 um2 focal spot and a flat panel amorphous silicon detector with 1024×1024 pixels. The microRT subsystem design utilizes a 320kVp orthovoltage source with dual focus spots (0.4×0.4 mm2 at 800W and 1×1 mm2 at 1800 W). The orthovoltage beam is collimated using two orthogonal jaws and exchangeable apertures. The resolution of the cone‐beam micro CT and the therapy beam spatial precision was determined using a numerical model. The microCT radiation dose, the orthovoltage source spectral output, and dose rate were evaluated using a mouse digital phantom (MOBY) and a pencil beam algorithm. Results: microCT reconstructed tomographic data with a resolution of 125 μm is achievable using 128 projections and a maximum radiation dose of 2cGy. Automatic animal positioning and handling is performed within a precision of 100 μm. The treatment beam can be aimed at different latitude and longitude angles in steps of 2 arc min. and translated at 50μm steps (x,y,z). The beam cross section can be modulated with submillimeter precision using steps of 50μm. A radiation dose rate of 40 Gy/min is delivered when the system is operated at an average half‐value layer of 4.6 mm Cu and a maximum beam penumbra of 0.35 mm. Conclusion: We designed and numerically evaluated the performance of a microRT system integrated with a microCT for conformal radiotherapy of murine animal models. This work supported in part by NIH grant CA108677.

AB - Purpose: To design a conformal small animal micro radiation therapy (microRT) instrument, consisting of a cone beam microCT subsystem for submillimeter low dose structural imaging and image guided radiotherapy and an orthovoltage conformal micro irradiator with high radiation dose rate for high throughput conformal irradiation. Method and Materials: The microCT subsystem is based on an 80kVp micro‐focus x‐ray source with 75×75 um2 focal spot and a flat panel amorphous silicon detector with 1024×1024 pixels. The microRT subsystem design utilizes a 320kVp orthovoltage source with dual focus spots (0.4×0.4 mm2 at 800W and 1×1 mm2 at 1800 W). The orthovoltage beam is collimated using two orthogonal jaws and exchangeable apertures. The resolution of the cone‐beam micro CT and the therapy beam spatial precision was determined using a numerical model. The microCT radiation dose, the orthovoltage source spectral output, and dose rate were evaluated using a mouse digital phantom (MOBY) and a pencil beam algorithm. Results: microCT reconstructed tomographic data with a resolution of 125 μm is achievable using 128 projections and a maximum radiation dose of 2cGy. Automatic animal positioning and handling is performed within a precision of 100 μm. The treatment beam can be aimed at different latitude and longitude angles in steps of 2 arc min. and translated at 50μm steps (x,y,z). The beam cross section can be modulated with submillimeter precision using steps of 50μm. A radiation dose rate of 40 Gy/min is delivered when the system is operated at an average half‐value layer of 4.6 mm Cu and a maximum beam penumbra of 0.35 mm. Conclusion: We designed and numerically evaluated the performance of a microRT system integrated with a microCT for conformal radiotherapy of murine animal models. This work supported in part by NIH grant CA108677.