The objective and subjective evaluation of low-frequency expansion in wide dynamic range compression hearing instruments

Kristy J. Lowery, Patrick Plyler

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

The effects of low-frequency expansion on the objective and subjective evaluation of four channel in-the-ear hearing instruments was investigated. Three expansion settings were programmed in each device: expansion off, expansion restricted to channel one, and expansion restricted to channels one and two. Objective evaluations were conducted in quiet (Connected Speech Test) and in noise (Hearing in Noise Test) with speech levels fixed at 40 dB SPL. Subjectively, each participant rated expansion satisfaction in quiet and listening to low-level speech in a sound-treated room then indicated the expansion condition preferred overall. Listeners performed significantly better in quiet and in noise for the Off and Channel 1 conditions than the Channels 1 and 2 condition; however, performance was similar between the Off and Channel 1 conditions. Expansion effects on listener satisfaction ratings depended on the listening environment. Overall, listeners preferred expansion in Channel 1 to expansion in Channels 1 and 2; however, preference was not significantly different between the Channel 1 and Off conditions. Results indicate restricting expansion to 1000 Hz overcomes speech-recognition deficits observed with expansion active across a broader spectrum without significantly reducing subjective benefit or preference.

Original languageEnglish (US)
Pages (from-to)641-652
Number of pages12
JournalJournal of the American Academy of Audiology
Volume18
Issue number8
DOIs
StatePublished - Sep 1 2007

Fingerprint

Hearing
Noise
Ear
Equipment and Supplies

All Science Journal Classification (ASJC) codes

  • Speech and Hearing

Cite this

@article{88b1c4bdca584ef29e275c3c60a026c0,
title = "The objective and subjective evaluation of low-frequency expansion in wide dynamic range compression hearing instruments",
abstract = "The effects of low-frequency expansion on the objective and subjective evaluation of four channel in-the-ear hearing instruments was investigated. Three expansion settings were programmed in each device: expansion off, expansion restricted to channel one, and expansion restricted to channels one and two. Objective evaluations were conducted in quiet (Connected Speech Test) and in noise (Hearing in Noise Test) with speech levels fixed at 40 dB SPL. Subjectively, each participant rated expansion satisfaction in quiet and listening to low-level speech in a sound-treated room then indicated the expansion condition preferred overall. Listeners performed significantly better in quiet and in noise for the Off and Channel 1 conditions than the Channels 1 and 2 condition; however, performance was similar between the Off and Channel 1 conditions. Expansion effects on listener satisfaction ratings depended on the listening environment. Overall, listeners preferred expansion in Channel 1 to expansion in Channels 1 and 2; however, preference was not significantly different between the Channel 1 and Off conditions. Results indicate restricting expansion to 1000 Hz overcomes speech-recognition deficits observed with expansion active across a broader spectrum without significantly reducing subjective benefit or preference.",
author = "Lowery, {Kristy J.} and Patrick Plyler",
year = "2007",
month = "9",
day = "1",
doi = "10.3766/jaaa.18.8.2",
language = "English (US)",
volume = "18",
pages = "641--652",
journal = "Journal of the American Academy of Audiology",
issn = "1050-0545",
publisher = "American Academy of Audiology",
number = "8",

}

TY - JOUR

T1 - The objective and subjective evaluation of low-frequency expansion in wide dynamic range compression hearing instruments

AU - Lowery, Kristy J.

AU - Plyler, Patrick

PY - 2007/9/1

Y1 - 2007/9/1

N2 - The effects of low-frequency expansion on the objective and subjective evaluation of four channel in-the-ear hearing instruments was investigated. Three expansion settings were programmed in each device: expansion off, expansion restricted to channel one, and expansion restricted to channels one and two. Objective evaluations were conducted in quiet (Connected Speech Test) and in noise (Hearing in Noise Test) with speech levels fixed at 40 dB SPL. Subjectively, each participant rated expansion satisfaction in quiet and listening to low-level speech in a sound-treated room then indicated the expansion condition preferred overall. Listeners performed significantly better in quiet and in noise for the Off and Channel 1 conditions than the Channels 1 and 2 condition; however, performance was similar between the Off and Channel 1 conditions. Expansion effects on listener satisfaction ratings depended on the listening environment. Overall, listeners preferred expansion in Channel 1 to expansion in Channels 1 and 2; however, preference was not significantly different between the Channel 1 and Off conditions. Results indicate restricting expansion to 1000 Hz overcomes speech-recognition deficits observed with expansion active across a broader spectrum without significantly reducing subjective benefit or preference.

AB - The effects of low-frequency expansion on the objective and subjective evaluation of four channel in-the-ear hearing instruments was investigated. Three expansion settings were programmed in each device: expansion off, expansion restricted to channel one, and expansion restricted to channels one and two. Objective evaluations were conducted in quiet (Connected Speech Test) and in noise (Hearing in Noise Test) with speech levels fixed at 40 dB SPL. Subjectively, each participant rated expansion satisfaction in quiet and listening to low-level speech in a sound-treated room then indicated the expansion condition preferred overall. Listeners performed significantly better in quiet and in noise for the Off and Channel 1 conditions than the Channels 1 and 2 condition; however, performance was similar between the Off and Channel 1 conditions. Expansion effects on listener satisfaction ratings depended on the listening environment. Overall, listeners preferred expansion in Channel 1 to expansion in Channels 1 and 2; however, preference was not significantly different between the Channel 1 and Off conditions. Results indicate restricting expansion to 1000 Hz overcomes speech-recognition deficits observed with expansion active across a broader spectrum without significantly reducing subjective benefit or preference.

UR - http://www.scopus.com/inward/record.url?scp=41449085191&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=41449085191&partnerID=8YFLogxK

U2 - 10.3766/jaaa.18.8.2

DO - 10.3766/jaaa.18.8.2

M3 - Article

VL - 18

SP - 641

EP - 652

JO - Journal of the American Academy of Audiology

JF - Journal of the American Academy of Audiology

SN - 1050-0545