Variations in pleural fluid WBC count and differential counts with different sample containers and different methods

Barrett D. Conner, Y. C.Gary Lee, Paul Branca, Jeffrey T. Rogers, R. Michael Rodriguez, Richard W. Light

Research output: Contribution to journalArticle

31 Citations (Scopus)

Abstract

Objective: To compare the results of pleural fluid analysis for WBC counts and differential cell counts as follows: (1) counting performed manually vs that performed by an automated cell counter; (2) cells collected in different types of specimen containers; and (3) cell counts performed at 4 and 24 h postthoracentesis. Methods: Twenty-eight pleural fluid samples were each collected in five different containers (ie, ethylenediaminetetraacetic acid (EDTA)-treated glass, citrate-treated glass, heparinized glass, plain glass, and plain plastic tubes). The WBC counts and differential cell counts were obtained manually (on the EDTA tube) and with an automated counter on all tubes within 4 h of collection, and again after 24 h of refrigeration. Results: There was a close correlation between the WBC counts obtained manually and those obtained with the automated counter from the pleural fluid samples collected in the EDTA tubes (r = 0.92). With the automated counter, the pleural fluid WBC counts were similar among the three tubes containing anticoagulants, but the counts obtained from the tubes without anticoagulants were significantly lower. The differential cell counts obtained manually and those obtained with the automated cell counter differed substantially. Although the percentage of lymphocytes was similar, the automated counter was inaccurate in differentiating neutrophils from large monocytes or mesothelial cells. The WBC counts obtained within 4 h of collection and 24 h after collection were virtually identical. Conclusions: The WBC counts obtained manually and with the automated counter from pleural fluid samples in EDTA tubes correlated very closely. The pleural fluid WBC count was lower if the pleural fluids had been collected in tubes without an anticoagulant. Automated WBC counts from pleural fluid specimens were inaccurate, possibly due to difficulty in separating neutrophils from monocyte/mesothelial cells. Refrigerated storage for up to 24 h had no significant effect on the total WBC count or on the WBC count differential regardless of the tube utilized.

Original languageEnglish (US)
Pages (from-to)1181-1187
Number of pages7
JournalChest
Volume123
Issue number4
DOIs
StatePublished - Apr 1 2003

Fingerprint

Edetic Acid
Glass
Cell Count
Anticoagulants
Monocytes
Neutrophils
Refrigeration
Citric Acid
Plastics
Lymphocytes

All Science Journal Classification (ASJC) codes

  • Pulmonary and Respiratory Medicine
  • Critical Care and Intensive Care Medicine
  • Cardiology and Cardiovascular Medicine

Cite this

Variations in pleural fluid WBC count and differential counts with different sample containers and different methods. / Conner, Barrett D.; Lee, Y. C.Gary; Branca, Paul; Rogers, Jeffrey T.; Rodriguez, R. Michael; Light, Richard W.

In: Chest, Vol. 123, No. 4, 01.04.2003, p. 1181-1187.

Research output: Contribution to journalArticle

Conner, Barrett D. ; Lee, Y. C.Gary ; Branca, Paul ; Rogers, Jeffrey T. ; Rodriguez, R. Michael ; Light, Richard W. / Variations in pleural fluid WBC count and differential counts with different sample containers and different methods. In: Chest. 2003 ; Vol. 123, No. 4. pp. 1181-1187.
@article{25ecaad6dd444e05b1ee4941302fa1b8,
title = "Variations in pleural fluid WBC count and differential counts with different sample containers and different methods",
abstract = "Objective: To compare the results of pleural fluid analysis for WBC counts and differential cell counts as follows: (1) counting performed manually vs that performed by an automated cell counter; (2) cells collected in different types of specimen containers; and (3) cell counts performed at 4 and 24 h postthoracentesis. Methods: Twenty-eight pleural fluid samples were each collected in five different containers (ie, ethylenediaminetetraacetic acid (EDTA)-treated glass, citrate-treated glass, heparinized glass, plain glass, and plain plastic tubes). The WBC counts and differential cell counts were obtained manually (on the EDTA tube) and with an automated counter on all tubes within 4 h of collection, and again after 24 h of refrigeration. Results: There was a close correlation between the WBC counts obtained manually and those obtained with the automated counter from the pleural fluid samples collected in the EDTA tubes (r = 0.92). With the automated counter, the pleural fluid WBC counts were similar among the three tubes containing anticoagulants, but the counts obtained from the tubes without anticoagulants were significantly lower. The differential cell counts obtained manually and those obtained with the automated cell counter differed substantially. Although the percentage of lymphocytes was similar, the automated counter was inaccurate in differentiating neutrophils from large monocytes or mesothelial cells. The WBC counts obtained within 4 h of collection and 24 h after collection were virtually identical. Conclusions: The WBC counts obtained manually and with the automated counter from pleural fluid samples in EDTA tubes correlated very closely. The pleural fluid WBC count was lower if the pleural fluids had been collected in tubes without an anticoagulant. Automated WBC counts from pleural fluid specimens were inaccurate, possibly due to difficulty in separating neutrophils from monocyte/mesothelial cells. Refrigerated storage for up to 24 h had no significant effect on the total WBC count or on the WBC count differential regardless of the tube utilized.",
author = "Conner, {Barrett D.} and Lee, {Y. C.Gary} and Paul Branca and Rogers, {Jeffrey T.} and Rodriguez, {R. Michael} and Light, {Richard W.}",
year = "2003",
month = "4",
day = "1",
doi = "10.1378/chest.123.4.1181",
language = "English (US)",
volume = "123",
pages = "1181--1187",
journal = "Chest",
issn = "0012-3692",
publisher = "American College of Chest Physicians",
number = "4",

}

TY - JOUR

T1 - Variations in pleural fluid WBC count and differential counts with different sample containers and different methods

AU - Conner, Barrett D.

AU - Lee, Y. C.Gary

AU - Branca, Paul

AU - Rogers, Jeffrey T.

AU - Rodriguez, R. Michael

AU - Light, Richard W.

PY - 2003/4/1

Y1 - 2003/4/1

N2 - Objective: To compare the results of pleural fluid analysis for WBC counts and differential cell counts as follows: (1) counting performed manually vs that performed by an automated cell counter; (2) cells collected in different types of specimen containers; and (3) cell counts performed at 4 and 24 h postthoracentesis. Methods: Twenty-eight pleural fluid samples were each collected in five different containers (ie, ethylenediaminetetraacetic acid (EDTA)-treated glass, citrate-treated glass, heparinized glass, plain glass, and plain plastic tubes). The WBC counts and differential cell counts were obtained manually (on the EDTA tube) and with an automated counter on all tubes within 4 h of collection, and again after 24 h of refrigeration. Results: There was a close correlation between the WBC counts obtained manually and those obtained with the automated counter from the pleural fluid samples collected in the EDTA tubes (r = 0.92). With the automated counter, the pleural fluid WBC counts were similar among the three tubes containing anticoagulants, but the counts obtained from the tubes without anticoagulants were significantly lower. The differential cell counts obtained manually and those obtained with the automated cell counter differed substantially. Although the percentage of lymphocytes was similar, the automated counter was inaccurate in differentiating neutrophils from large monocytes or mesothelial cells. The WBC counts obtained within 4 h of collection and 24 h after collection were virtually identical. Conclusions: The WBC counts obtained manually and with the automated counter from pleural fluid samples in EDTA tubes correlated very closely. The pleural fluid WBC count was lower if the pleural fluids had been collected in tubes without an anticoagulant. Automated WBC counts from pleural fluid specimens were inaccurate, possibly due to difficulty in separating neutrophils from monocyte/mesothelial cells. Refrigerated storage for up to 24 h had no significant effect on the total WBC count or on the WBC count differential regardless of the tube utilized.

AB - Objective: To compare the results of pleural fluid analysis for WBC counts and differential cell counts as follows: (1) counting performed manually vs that performed by an automated cell counter; (2) cells collected in different types of specimen containers; and (3) cell counts performed at 4 and 24 h postthoracentesis. Methods: Twenty-eight pleural fluid samples were each collected in five different containers (ie, ethylenediaminetetraacetic acid (EDTA)-treated glass, citrate-treated glass, heparinized glass, plain glass, and plain plastic tubes). The WBC counts and differential cell counts were obtained manually (on the EDTA tube) and with an automated counter on all tubes within 4 h of collection, and again after 24 h of refrigeration. Results: There was a close correlation between the WBC counts obtained manually and those obtained with the automated counter from the pleural fluid samples collected in the EDTA tubes (r = 0.92). With the automated counter, the pleural fluid WBC counts were similar among the three tubes containing anticoagulants, but the counts obtained from the tubes without anticoagulants were significantly lower. The differential cell counts obtained manually and those obtained with the automated cell counter differed substantially. Although the percentage of lymphocytes was similar, the automated counter was inaccurate in differentiating neutrophils from large monocytes or mesothelial cells. The WBC counts obtained within 4 h of collection and 24 h after collection were virtually identical. Conclusions: The WBC counts obtained manually and with the automated counter from pleural fluid samples in EDTA tubes correlated very closely. The pleural fluid WBC count was lower if the pleural fluids had been collected in tubes without an anticoagulant. Automated WBC counts from pleural fluid specimens were inaccurate, possibly due to difficulty in separating neutrophils from monocyte/mesothelial cells. Refrigerated storage for up to 24 h had no significant effect on the total WBC count or on the WBC count differential regardless of the tube utilized.

UR - http://www.scopus.com/inward/record.url?scp=0037390459&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0037390459&partnerID=8YFLogxK

U2 - 10.1378/chest.123.4.1181

DO - 10.1378/chest.123.4.1181

M3 - Article

C2 - 12684309

AN - SCOPUS:0037390459

VL - 123

SP - 1181

EP - 1187

JO - Chest

JF - Chest

SN - 0012-3692

IS - 4

ER -